microGC 490 (obsolete)

The Agilent 490 Micro GC is a rugged, compact, laboratory quality gas chromatograph that delivers information you need, wherever you need it. The fifth generation Agilent 490 Micro GC is designed to provide maximum flexibility and ease-of use. You can customize with options including column modules and optimized sample conditioning to achieve greater sensitivity and performance. 

Your Agilent 490 Micro GC arrives pre-configured, factory tested for your analysis and ready to go. If your needs change, you can reconfigure in the field for new applications in just minutes with our user installable plug and play GC channels. The small size and weight of the Agilent 490 Micro GC, make it easy to carry between analysis sites in the field, such as perimeter monitoring, multiple drilling locations, and at natural gas pipelines and metering stations.

 

- Fast, precise gas analysis in seconds, not minutes with narrow-bore capillary, PLOT or micro-packed columns to bring improved product quality and more exact product valuation to your field applications

- Flexibility to run in the lab, at-line or on-line with up to four independent analytical GC channels, each with pneumatics, injector, column and detector. Quickly reconfigure for any application

- Increase Monitoring Frequency through micro-electronic gas control with optional time-programmable backflush so you can inject samples while eliminating components that could foul the column and reduce lifetime

- Quick and easy start up so you can get the results and ruggedness you demand in the lab and in the field. You can move from location to location and get results in minutes.

- Powerful, yet easy-to-use Agilent OL EZChrom, EZChrom and EZChrom Elite software gives you comprehensive system control, data acquisition and report generation. Or control the Micro GC from virtually anywhere using a PC and internet, with industry standard protocols such as TCP/IP.

 Configuration and Control

 Channel

 From 1 to 4 channels

 Channel Control

 Independent control of each analytical GC channel

 Control Type

 Pneumatic

 Column

 Temperature

 Up to 180 ºC

 Control T

 Isothermal

 Detector

 Type

 Micro-machined Thermal Conductivity Detector (TDC)

 Channel

 Dual-channel (sample and reference flow)

 Volume Internal

 200 nL per channel

 Filaments

 4

Limits Detection, TCD

 WCOT 1ppm. Micro-packed/PLOT 10ppm

 Injector

 Control Type

 Micro-machined injector with no moving parts

 Injection Volume

 1 µl a 10 µl

 Temperature

 Optional heated injector, up to 110 ºC, including heated sample transfer line

 Backflush

 Optional backflush capability for each module

 Repeatability

 RSD

 < 0.5% RSD for propane at 1 mol % level for WCOT columns at constant temperature and pressure

 Carrier Gas

 Input

 He, H2, N2, or Ar, 550 +/- 10 kPA (80+/- 1.5 psi) input

 Gas Type

 Up to 2 different types of carrier gases can be used in one instrument

 Connection

 Inlet connection, 3.2 mm (1/8 in.) stainless Steel compression fitting

 Sampling

 Sample Inlet

 1.6 mm (1/16 in.) Stainless Steel Valco fitting with replaceable 5 µm SS filter

 Conditions

 Sample conditions, non-condensing gas of 0 ºC to 110 ºC.

 Maximum sample inlet pressure 

 100 kPA (14.5 psi)

 Software

 Selectable sample pump or continuous flow

 Options

 Optional sample inlet can be installed in front or back

 Communication

 Analog Input

 6 signals (0-10V)

 Type

 LAN (TCP/IP)

 RS-232 Connection

 For control of a stream selection valve

 Webserver

 Display instrument status on standard Internet browser

 Environmental Requirements

 Humidity (Relative)

 0 % to 95 % (non-condensing)

 Temperature

 0 ºC to 50 ºC

 Latitude

 Certified up to 2000 m above sea level

 Power Requirements

 Power

 Main power, 90-130 Vac or 180-260 Vac, 50-60 Hz

 Output

 Output, 12 VDC, maximum 130 W