DVLS3 Simply, Smart Sensor

The DVLS3 Simply, Smart Sensor is installed in the GC oven for continuous monitoring of the hydrogen concentrations in the oven air. The LCD screen of the external controller displays the measured gas concentration. If the sensor detects a user defined hydrogen concentration, typically between 25% and 50% LEL (equal to 1%-2% by Vol. hydrogen), it will ensure the safety of the laboratory staff by several means:

  1. The LCD screen of the sensor controller starts flashing and the controller transmits an acoustic signal.
  2. The sensor software alerts the lab staff by showing a red status on the dashboard and sending an alarm message by e-mail, text (SMS), Growl or Prowl.
  3. The sensor controller automatically switches the carrier gas to an inert gas and sends an automatic stop signal to the GC.

 

Due to the unique design of the DVLS3 Simply, Smart Sensor you can combine up to four sensors in one controller. For example use four sensors for several GC’s or connect several sensors to one GC.

The sensor is compatible with all GC brands. Next to the hydrogen leak detection Da Vinci Laboratory Solutions offers multiple sensors dedicated to the detection of:

Hydrogen (H2)

- Temperature

- Barometric Pressure

- Level (liquid) weight

DVLS3 Simply, Smart Sensor benefits:

- Safe technique to constantly monitor hydrogen leaks

- Automatic alarm transmission through:

         1. Optical and acoustic alarms

         2. E-mail messages

         3. Text (SMS) messages

         4. Growl

         5. Prowl

- Dedicated software controls the alarm settings per sensor & user.

- La función de registro de auditoría crea un registro de las alarmas detectadas y de los ajustes.

- Audit trail function creates a record of the reported alarms and settings.

- Automatic carrier gas switching from Hydrogen to Nitrogen.

- Leak detection indicates analysis is compromised.

- Automatic stop signal sent to the GC.

- Easy calibration and installation.

- Compatible with all GC brands.

- Versatile design supports up to 4 sensors.

- Versatile multi sensor offers GC labs a safe solution to change the GC carrier gas to hydrogen.

 

 Number of sensors

 Max 4 sensors individually controlled

 Detection range

 0-50% LEL ( 0 – 2.0% by vol.Hydrogen)

 Installation

 Gas sensor integrated in GC oven, carrier gas switch and LCD display are installed in an external controller 

 Sensor

 Catalytic pellistor gas sensor, linear range of 0-2% Hydrogen (0-50% LEL)

 Alarms

 User defined alarm level optical and/or acoustic alarm up to 50% LEL

 Instrument readings

 Provide real time sensor readings with alarm levels, channel status

 Menu

 User menu for configuration

 Valve

 High pressure 3 way solenoid valve

 Oven operating temperature 

 Up to 450°C 24VDC power adapter

 Dimensions

 11 cm (W), 5,5 cm (H), 14 cm (D)