The Disulfides Analyzer has been developed for the analysis of (di-) sulfides such as (di-)glutathione, cystine, cysteine, homocystine, and homocysteine. The method applies a Flexcell with the new Magic Diamond working electrode for detection of mono and disulfides. Detection limits are below 0.1 µmol/L for most (di-)sulfides. In an alternative setup a dual flowcell configuration is used with about 10 times better detection limit. The method uses an additional Flexcell for pre-reduction of di-sulfides, a second cell is used for detection.
- Both mono- and disulfides in one run
- Single flow cell with diamond electrode, or
- Dual flow cell for selectivity with gold electrode
- Detection limits 2-10 (gold) or 25-100 nmol/L (diamond)
The chromatogram shows the analysis of 2 µL, 1 µmol/L cystine (1), cysteine (2), glutathione (3), homocysteine (4) and diglutathione (5). If only the mono-sulfides are of interest, the Flexcell can be equipped with a Gold working electrode. This offers additional selectivity, and the lower working potential results in a better detection limit as well (10 nmol/L).