NG POLLUCE

POLLUCE is designed to satisfy the needs of all LCMS brand on the market. It makes use of Pressure Swing Adsorption technology (PSA), through the use of specific selected molecular sieves (CMS), enable to obtain extremely pure nitrogen, this technologies reduced risk of gas contamination and phthalates-free.

The generator could be equipped with a built-in oil less compressor with inverter technology. Inverter electronically control the electrical voltage, current and frequency of the compressor motor. It receives information from sensors monitoring operating conditions, and adjust the revolution speed of the compressor, which directly regulates air pressure output. Optimum control of operation frequency results in eliminating the consumption of excessive electricity and providing the most comfortable conditions to increase the compressor life-time. Thanks to the CPU with touch screen display, the continuous monitoring of operational status is guaranteed.

Main application

  • LCMS
  • Solvent evaporation

Main advantages 

  • PSA Technology
  • Easy use and installation (plug and play)
  • Low noise
  • Generates nitrogen on demand with exclusive electronic flow control
  • Low energy exclusive compressor inverter controlled
  • Designed to run 24 hours a day
  • Pressure up to 8 bar
  • CH4 oven scrubber (option)

Technical specifications

Models: NG POLLUCE

40

40C

100

N2 outlet

Flow rate (max)

40 L/min

40 l/min

100 L/min

Outlet pressure (Max)

Air Inlet pressure–2 bar

Up to 8 bars

Air Inlet pressure–2 bar

Nitrogen purity

Up to 99%

Outlet Dew-point

<-50°C

Outlet particulate

0.01 micron

Air inlet

 

Air inlet requirement

Particulate: <1 micron

Dewpoint: < -20°C (- 4°F)

Oil: <0.01 mg/m³

Flow rate (Max)

100 L/min

250 L/min

Supply pressure (min; max)

7 bars (95 psi)/ 10 bars (145 psi)

Recommended Temperature

< 30°C (86°F)

Communication

LCD with touch screen

Standard

RS485

Standard

RS232

Standard

USB

Standard

Digital I/O

Standard

General data

Supply rating

100-240Vac  50/60Hz

Single phase 220-40Vac 50/60Hz

100-240Vac  50/60Hz

Connection type

IEC320-C13

Cable, 3x2.5mm²

IEC320-C13

Installation power (max)

280W

2.2KW

280W

Fuse rating (5x20mm)

4A - 250VAC

-

4A - 250VAC

Net weight

<160Kg

<300Kg

<220Kg

Dimensions (W x D x H)

42 x 40 x 160 cm

42 x 90 x 160 cm

42 x 80 x 140 cm

Connections

Outlet port

¼" BSPP female

Inlet port

¼" BSPP female

-

¼" BSPP female

Drain port

Quick fitting for tube 8mm

¼" BSPP female

Quick fitting for tube 8mm

Operating/storage conditions

Working Temperature

5-40°C (41-104°F)

Storage Temperature

1-50°C (34-122°F)

Humidity (max)

70% (non-condensing)

Noise

<25dB(A)

<65dB(A)

<25dB(A)

IP rating

IP20

Pollution degree rating

2 (with no aromatic compounds)

Altitude

< 2000m

 

Ordering codes

Gas Generators

6920.60.040 NG POLLUCE 40

6920.60.040C NG POLLUCE 40C

6920.60.100 NG POLLUCE 100

Options

6920.61.001 CH4 scrubber for NG POLLUCE (only for 40/ 40C)

Consumable

6920.61.002 Kit service 8000 hours for NG POLLUCE