5975T LTM GC/MS
When you need fast and accurate results in the field, the Agilent 5975T LTM GC/MSD can deliver the same high performance, reliability and quality of analysis as our bench top 5975C series GC/MSD. The 5975T system is the first transportable commercial GC/MSD system that can provide credible results on the ground without any lab-based analysis required.
The system is a third smaller and consumes about half the power of our bench top option, enabling operators to quickly and efficiently provide the credible results needed to make vital health and safety decisions. The 5975T system utilizes Agilent’s proprietary LTM technology for direct, rapid heating and cooling, enabling faster on-site analysis across a wide mass range. Check our web www.yourmobilelab.com for the integration of these systems in our mobile units.
-
1.8u to 1050u mass range for the greatest range of applications
-
Transportable and highly reliable for quick response in the field.
-
Proprietary LTM technology allows for direct, rapid heating (up to 1200°C/min) and cooling for faster on-site analysis
-
Gold true hyperbolic quartz quadrupole provides unsurpassed performance and reliability for on-site measurement
-
Classical EI spectra using quadrupole technology delivers NIST searchable spectra of unknowns
-
7667A Mini Thermal Desorber lets you perform fast, confident gas sample introduction in the field
-
Thermal Separation Probe(TSP) – A fast and easy alternative MS probe for analyzing solid, liquid, and slurry samples in the field
-
Handheld Capillary Trap Gas Sampler (CTS) puts air sampling at your fingertips – no matter where your analysis takes you
-
Agilent’s 7693 Series and 7650A Automatic Liquid Sampler (ALS) deliver the fastest injection times of any GC autosampler
-
Excellent value for money with one system for both lab and field use
Mass Spectrometer |
|
Mode |
EI (Electron Impact Ionization) |
Ion source type |
Noncoated inert EI source |
Ionization energy |
5-241.5 eV |
Ionization current |
0-315 uA |
Transfer line temperature |
100-350 ºC |
Ion source temperature |
150-350 ºC |
Quadrupole temperature |
106-200 ºC |
Filaments |
Dual for EI |
Mass filter |
Monolithic hyperbolic quadrupole |
Scan rate (electronic) |
Up to 12500 u/s |
Detector |
Triple Axis Detector with long life EM |
Mass range |
1,6-1050 u |
Pumping System |
70 L/s turbomolecular pump with 3.6 m3/hr standard dry pump |
Total flow |
2 mL/min (standard turbo) |
Instrument control |
MSD Chemstation Data System, LTM GC remote control panel software |
LTM Technology |
|
Column temperature range |
Ambient +8 ºC to 350 ºC |
Ramps |
20/21 with negative ramps allowed |
Max heating rate |
1200 ºC/min (depends on column length and temperature) |
Autosampler |
Agilent 7693A (G4513A only), CombiPAL, Headspace, Purge&Trap, Thermal desorber |
Injector |
S/SL Standard |
Inlet max temperature |
350 ºC |
Carrier gases |
Helium or hydrogen |
EPC (Electronic pneumatic control) |
Installed in injectors and detectors. |
Zones with independent temperature control |
5 (2 injectors, 2 detectors, 1 auxiliary) |
Physical Requirements |
|
Dimensions (GC/MSD) |
65 cm (w), 60,8 cm (d), 49,1 cm (h) (without accessories) |
Weight (GC/MSD) |
56,5 kg (without pump) and 67,5 Kg (with pump) |