The CDS 9350 is the latest real time air monitoring system.  It has been designed to deal with the more difficult environments that have to date, not been achievable.  This includes the ability to collect and analyze compounds with boiling points down to -50º C, without cryogen trapping or cryofocusing, and collect analytes with boiling points above 300º C with a sampling line and all heated zones capable of maintaining temperatures of up to 350º C.  

The CDS 9350 has fast flow capabilities for sample collection of up to 1.5 LPM, rapid heat and cool down times for both sorbent tubes and focus trap without the need for cryo focusing.  This allows for much quicker cycle times, from collection to completion of analytical analysis.

 

- Compatible with all GC and GC/MS makes and models.

- Dual sorbent tubes for continuous sampling

- Single focus trap

- Desorption temperatures up to 375°C

- Valve and transfer Line temperatures up to 350°C

- Heating rates: 1000°C/min

- Sorbent tubes: Standard 6 mm * 4 ½”, Fast Flow 10 mm * 4 ½”(6 mm ends)