Sensores de hidrógeno Ambientales
View all
Environment H2 sensor kit 6940.00.010 & 6940.00.020
Environment H2 sensor kit 6940.00.010 & 6940.00.020
Preview

This H2-SENSOR Kit is a security system that constantly detects the amount of Hydrogen present in the working environment of the GC and blocks the Hydrogen generation if it detects an amount greater than the safety threshold.

The 6940.00.055 controller reads the percentage of Hydrogen measured by the sensor 6940.00.060 and, if the value exceeds the threshold alarm, the Controller deactivates the solenoid valve in the 6940.00.056 block.

The alarm can be signaled to a PC and to a remote device.


El sensor puede conectarse directamente con los generadores de hidrógeno LNI. Cuando la concentración de hidrógeno detectada supera el umbral de seguridad, el generador de hidrógeno detiene automáticamente la producción y se activan alarmas auditivas y visuales.


Especificaciones técnicas SENSOR AMBIENTAL

Modelos: 6940.00.050
6940.00.055
Voltaje de alimentación 12 VDC +/- 5%
Corriente de alimentación0.8 A use with 6940.00.056
Valve outputOpen collector
12VDC/500mA
Ambient temperature5-35°C (41-95°F)
Ambient Humidity (max, non condensing)80% at 25°C (77°F)
IP ratingIP20
6940.00.056
Hydrogen Inlet pressure1 to 9 bar
Inert Inlet pressure1 to 10 bar
Pressure regulator1 to 9 bar

 

6940.00.060

Sensores de hidrógeno Para Cromatógrafos de Gases
View all
H2 sensor kit for GC oven
H2 sensor kit for GC oven
Preview

Technical specifications Kit de sensor de H₂ para horno de GC

 6940.00.050                      

 

    6940.00.055
Power Supply Voltage12 VDC +/- 5%
Power Supply Current0.8 A use with OPT.H2.SENS.VB.V2
Valve outputOpen collector

12VDC/500mA

Ambient temperature5-35°C (41-95°F)
Ambient Humidity (max, non condensing)80% at 25°C (77°F)
IP ratingIP20

 

 6940.00.056
Hydrogen Inlet pressure1 to 9 bar
Inert Inlet pressure1 to 10 bar
Pressure regulator1 to 9 bar

 

               6940.00.070

 

Hydrogen Sensitivity Range0.01% to 1.00% by hydrogen volume (25% LEL, Lower Explosion Limit)
Accuracy± (5.0% of F.S. + 1000ppm)
Typical Response Timeless than 20 sec
Power Supply VoltageFrom 9VDC to 28VDC
Communication portRS-485
ProtocolMODBUS-RTU
Accuracy checking interval1 year (suggested 180 days)
Hydrogen Sensor TechnologyDioxide semiconductor
Product Life Expectancy10 years
Body temperature range-10 °C to 40 °C