TD100-xr Markes International

TD100-xr™ is a high-throughput, automated thermal desorption system for the rapid and unattended processing of up to 100 sample sorbent tubes in a single sequence.

In contrast to other thermal desorption units, TD100-xr quantitatively re-collects samples for re-analysis or storage, enhancing sample security.

TD100-xr also features an electrically cooled cold trap, dispensing with the expense and inconvenience of liquid cryogend trap, dispensing with the expense and inconvenience of liquid cryogen.

Product benefits

• Enhance throughput – capacity for up to 100 tubes and compatibility with the widest range of TD applications allow you to maximise productivity, e.g. unattended operation all weekend.

• Peace of mind – with quantitative sample re-collection for repeat analysis, TD100-xr offers simple method/data validation and overcomes the ‘one-shot’ limitation of other systems.

• Enhance productivity with the world’s most versatile thermal desorber – TD100-xr offers unmatched analytical performance for the widest range of TD applications.

• Reduce running costs by avoiding the need for cryogen.

• Security – tubes remain capped at all times, so avoiding sample contamination and/or analyte loss.

• Reduce errors with the TubeTAG™ sample informatics system.

• Be confident – you’re using a desorber that complies with key standards such as US EPA Method TO-17.

Features

• Automated, cryogen-free, unattended operation for up to 100 sample tubes.

• Confidence in results through quantitative sample re-collection of split flows.

• Inert sample paths and extended temperatures allow quantitative recovery of C2 to C44, including reactive and thermally labile species... from percent to sub-ppt concentrations.

• Method compliance aided by leak-testing, water management and addition of internal standard.

• Enhanced traceability of samples using barcodes and RFID TubeTAGs.

• The short, heated transfer line allows TD100-xr to be installed on all major makes of GC and GC–MS.

• High-precision parts result in increased robustness.

 

Desorption tubes
 The automated desorber must be compatible with industry standard 3.5-inch (89 mm) x ¼-inch (6.4 mm)  O.D. tubes for compliance with standard methods e.g. ASTM D 6196, EN ISO 16017, ISO 16000-6, EN  14662-4, US EPA Method TO-17, NIOSH 2549 and the UK MDHS series.
 The automated desorber must be compatible with RFID tagged tubes
 The tubes must be compatible with standard Swagelok-type ¼-inch long-term brass storage caps fitted with  combined PTFE ferrules as specified in standard methods.
 Sample tubes must be compatible with axial diffusive sampling as per standard methods (EN ISO 16017-2,  EN 14662-4 and ASTM D 6196).
 Tubes should be orientated horizontally throughout TD operation to prevent particles dropping into tube  seals and to prevent samples shifting within the tube during direct desorption of materials.
 Carrier gas must only be allowed to pass through the middle of the tube and not around the outside of the  tube during the analysis process – to minimise contamination. 

 

Focusing trap
 The system must feature 2-stage desorption using an electrically-cooled packed cold trap capable of cooling  to sub-zero temperatures as specified in US EPA Method TO-17 and other standards.
 It should be possible to pack the electrically-cooled cold trap with up to four sorbent beds to a total length of  60 mm to match the sorbent bed length found in industry standard sample tubes. 
 Desorption of the focusing trap must take place in a backflush mode i.e. analytes must enter the trap from  one end and be desorbed from the same end to facilitate simultaneous analysis of volatiles and semi-  volatiles. 
 The system should offer rapid heating (>60°C/sec) of the focusing trap to optimise chromatographic  resolution and analytical sensitivity. 
 Optional, slower, programmed heating options should also be available for the focusing trap to ensure  compatibility with very reactive components
 The focusing trap must feature a narrow / restricted I.D. (<1 mm I.D.) at the inlet / outlet end to the cold  trap to optimise linear gas velocity and minimise band width after trap desorption.
 The dimensions, sorbent mass and heating rate of the focusing trap plus the design of the sample flow path  downstream of the focusing trap should facilitate splitless operation even with high resolution  capillary GC / GC-MS operating at flow rates of 2 mL/min or below.
 The flow of dry gas required for purging the cold trap area to prevent build-up of ice during system  operation, should be less than 100 ml/min to minimise gas consumption and operator intervention.

 

Sample Flow Path
 The desorber must feature a totally inert flow path (constructed, for example, using SilcosteelÒ, quartz or  fused silica components) the flow path must contain no uncoated metal fittings.
 The sample flow path inside the desorber must feature uniform heating and minimal volume / length / I.D.,  particularly post focusing trap, to ensure high linear gas velocity and no band dispersion as components are  transferred to the capillary column or fused silica retention gap.
 It must be possible to set the flow path at temperatures below 100°C to prevent degradation of heat  sensitive compounds.  It must also be possible to set the flow path at temperatures which allow quantitative  recovery of low volatility compounds, including n-C40
 After primary (tube) desorption, the desorber must isolate the cooling sample tube from the flow path to  prevent contaminants desorbing late from the primary tube from passing into the focusing trap after  secondary (trap) desorption has begun.  Isolation of the primary tube should involve some sort of  appropriate heated, inert valve
 At no stage during sample analysis must the sample flow path be opened to atmosphere or air (and  associated contaminants) will get into the system
 Once the desorber has initiated a run, all parts of the desorption sequence – leak testing, purging, tube  desorption, trap desorption and starting the GC run – should be automatic, to allow unattended operation.
 The desorber must connect directly to the analytical column or to a heated uncoated, deactivated fused-silica  capillary retention gap, without passing through a GC injector

 

Sample split capabilities
 It should be possible to split both during primary (tube) desorption (inlet splitting) and secondary (trap)  desorption (outlet splitting) – in order to obtain overall split ratios in excess of 1000:1 for analysis of high  concentration samples.
 The flow path to the sample split point(s) must be inert and uniformly heated
 The system must allow the user to select for the sample split flow to be on during instrument standby to  minimise air ingress for MS and ECD installations.
 The system must allow manual, quantitative re-collection of both the inlet and outlet split effluent on a  single re-collection tube such that the entire unanalysed portion of the sample, from any single or double  split method, is re-collected. The re-collected sample is needed for repeat analysis, sample archiving and  method / data validation in compliance with ASTM 6196.
 There must be easy access to the split re-collection tube without need to tighten and undo screw fittings.  Software must also allow carrier gas to be isolated from the split line such that access to the re-collection  tube does not cause a temporary leak.
 It must be possible to automate split re-collection – this upgrade should be possible at any point during the  lifetime of the system

 

Analytical sequence
 As required by relevant standard methods such as US EPA Method TO-17, the analytical sequence must  include an ambient temperature, no-flow leak test of each tube prior to analysis, to ensure data integrity.
 Any tube which fails the leak test must remain intact and unanalysed, maintaining sample integrity until the  fault is diagnosed and the sample can be successfully analysed.
 The analytical sequence must also include an ambient temperature carrier gas purge of each tube to  eliminate O2 prior to analysis.  Any air purged from the system must be directed away from the analytical  GC(MS) system. This is in order to minimise interference – e.g. column oxidation or high air / water  background on MS.
 The system must offer the option of an elevated temperature tube purge during standard 2-stage tube  desorption for selective elimination of water and unwanted volatiles prior to analysis.
 The desorption oven should heat from near ambient during primary (tube) desorption to minimise risk of  split discrimination from samples with high solvent or water content
 The system should facilitate the sequencing of multiple desorptions of a single tube at different temperatures  for automating the TD method development process

 

Autosampler operation
 The automated TD should be compatible with RFID-tagged or untagged tubes and include    automatic tag read/write to facilitate tube tracking and automate the process of inserting sample/tube    specific information into the automation sequence
 For maximum productivity (e.g. operation throughout an entire weekend), the system should have capacity    for up to 100 industry standard sample tubes.
 For maximum productivity it must be possible to operate the automated desorber in ‘overlap’ mode – i.e.    for the system to begin pre-desorption tests and primary desorption of a subsequent sample, while the    chromatographic cycle of the previous sample is still in progress.
 The system must offer the option of dry purging tubes with carrier gas, in the sampling direction and at    ambient temperature, as part of the automatic analytical sequence in compliance with US EPA Method TO-17
 Tubes must be properly sealed before and after sampling with inert, non-emitting caps. Neither septa seals    nor abutting the tube ends against a flat surface is a suitable sealing mechanism
 The tube caps used on the TD autosampler must not use PTFE-coated o-rings as seals as these   have been shown to leak causing loss of analyte and possible ingress of contaminants
 To minimise risk of mechanical failure the instrument should not be required to remove or replace caps    during analysis.
 Every tube should be subjected to the integrated ambient temperature, no-flow leak test before desorption    and should not be analysed if any leak is detected.  Failed tubes should be left intact and logged in system    memory awaiting user intervention.  After a tube has failed the leak test, the system should continue to test    and analyse subsequent tubes.
 If a tube fails the leak test during an automatic sequence, a GC run should be initiated to keep the desorber    sequence in-synch with that of the GC(-MS.)  The system should stop if a more serious error (i.e. one    requiring user intervention to the system itself) occurs.
 The automated desorber must offer the option of automatic addition of gas-phase internal standard to each    sorbent tube in the sampling direction. 
 Standard addition should only occur after leak testing and before any other stage of operation in compliance    with the recommendations of key standard methods such as US EPA Method TO-17
 The system should also allow gas-phase standard to be introduced to the sampling end of blank tubes after    leak testing, but without subjecting them to the desorption process.  This will enable blank tubes to be pre-    loaded with internal standard before being used for field monitoring 

 

Automation of split re-collection
 Automatic re-collection of the outlet (trap) split flow from up to 50 sample tubes must be possible  without using the original (primary) sample tube for the re-collection process (NB Using the original  sample tube to re-collect the split flow would render any validation work meaningless). PATENT GB2395785
 Systems offering automatic re-collection of outlet split flow on a single TD autosampler should also offer          manual re-collection of inlet (tube) and outlet (trap) split flow in order to facilitate validation of inlet or  
 double split methods
 It should also be possible to upgrade to automated re-collection at any time