Stratum
The Stratum PTC builds on eight previous generations, making us the industry’s cornerstone of technique and support. The Stratum PTC uses components that are absolutely essential to perform a successful Purge and Trap analysis; those being a mass flow controller (MFC), superior trapping technology, and inert sample pathway.
The heart and soul of our Purge and Trap System is the Teledyne Hastings Mass Flow Controller. The MFC delivers the purge gas with unmatched precision and accuracy. The most advantageous feature of utilizing the MFC is the ability to instantaneously change the gas flows and pressures at any time during the Purge and Trap process.
Including the The StratTrap system, is capable of adsorbing the VOCs of interest while minimizing the collection of unwanted compounds, such as water. The ultra-fast trap heater ensures that the StratTrap is desorbed rapidly and the VOCs are delivered to the GC system quickly.
One important key to minimize carryover is the Stratum PTC’s inert sample pathway. The tubing and fittings have been treated to provide unprecedented low carryover which permits the enduser to calibrate with more confidence.
Automation |
|
Cycle Time | The cycle time for the unit is 17 minutes when using an 11 minute purge time. This time also includes desorb, bake, and cool down for the Stratum PTC only and assumes ambient lab temperature. (20-22°C). |
Capacity | 80 positions for 40 mL vials |
Trap Furnace | Ambient to 350°C cools from 250°C to 40°C in 90 seconds or less at ambient lab temperatures (20-22°C). |
6-port switching valve | Ambient to 300°C actuated at 24 Volts (D.C.). |
External Transfer Line | Ambient to 300°C. |
Sample Mount | Ambient to 100°C. |
Condenser | Ambient to 250°C. |
Sample Heater (optional) | Ambient to 90°C. |
Sample Pathway | All tubing and related fittings use Siltek® coating. |
Gas Requirements | 99.999% Helium or Nitrogen |
Electronic Mass Flow Controller |
Device is capable of controlling flow rates between 5mL/min to 500mL/min. Each mode is independently controlled. Device also capable of recording pressures for sample logging and automatic leak checking |
Unit Dimensions | Height: 19.75 inches (50.17cm) Width: 8 inches (20.32cm) Depth 18.5 inches (46.99cm) |
Operating System | PC using Windows® XP or greater |
Software | VOC TekLinkTM interfaced via an RS-232 connection. |
Operating Conditions | The system is capable of operating in Lab Temperatures between 10-30°C and humidity levels between 10-90% |
Corrosion | The front cover is corrosion resistant to waters within a pH range of 1-10 |
Voltages | 100/115VAC 50/60Hz 10 amps, 1150 watts 220/240VAC 50/60Hz, 5 amps, 1150 watts |
Weight | 32lbs |