GDA-S

Monitoring of Critical Infrastructures. Detector for hazardous gases and chemical agents. The Gas Detector Array Stationary version is a continuously operated chemical agent detection system. It is used to supervise sensitive public structures, buildings and workplace areas. The benefit using GDA technology is, that not only chemical warfare agents (CWAs) are selectively supervised but also the whole range of hazardous and less hazardous volatile compounds can be supervised since GDA technology offers the possibility to detect a very broad range of compounds in the gaseous phase.

The GDA-S version has been developed on the basic idea of combining several detection principles in order to achieve: a broad detection range and thus giving a high level of safety and improved specifity through combined sensor responses can be used for library comparison. The non-stop monitoring capability and broad detection range of the GDA-S provide a high level of security.

 

 

The changes of the specifications made for the stationary GDA are:

• Fail safe flow system (e. g. pumps allowing long term continuous operation, redundancy included)

• Maintenance interval designed to be 1 year

• Connectivity – personal computer connectivity offers all common kind of data interfacing

• Adaptable library system

The detector is rugged, reliable and dependable, even operating in adverse environments, but quickly and easily maintained at yearly service intervals.

Features:

• Detection and identification of all the main hazardous gases and chemical warfare agents within seconds

• Hybrid Sensor Array: Unique combination of different detectors (IMS, PID, EC, MOS)

• Safe alarming concept

• Alarm and communication interface

• 24/7 operation / data stored

• Internal sensor protection system

• Easy to install

• Database is expandable

• Outdoor operation

Sampling

 System

continuous vapour sampling through internal pumps, internal sample dilution system

 Recovery time

less than typically 5 min

 Measurement time

seconds to less than 1 minute (depending on the compound)

Operation Principles

 Detection principles

Orthogonal technology for improved interferent rejection:

- Ion Mobility Spectrometer (Ni63 ion source, positive and negative mode)

- Photo Ionization Detector (10.6 eV)

- Electrochemical Cell

- 2 Metal Oxide Sensors

 Modes of operation

GDA mode for hazardous compounds and chemical warfare agents

 Agents detected

 nerve, blister, blood & choking agents, toxic industrial chemicals, data base is expandable 

 Identification

based on pattern recognition methods, individual alarm thresholds are possible

Environment Requirements

 Temperature

typical: 0ºC to +50ºC

 Humidity (relative)

5 % to 95 %, non-condensing

Power Requirements

 Main power

30 W, powered by power supply of 100 – 240 Volt

 Battery back-up

Operation on Backup Battery.

Battery to be recharged by internal charging circuit (UPS)

 Communication

 Computer interface 

Internal IPC

Devise Control / Data Handling

 Operating system

Windows XP, Vista, Windows 7

 Software

WinMuster GDA-S

 Safety Class

Compliant to EN50270 / 1999 / type 1 & 2 device