325 Dual Wavelength UV-Vis Detector
The Agilent 325 Dual Wavelength UV-Vis Detector delivers superb sensitivity for detection in the UV and Visible light ranges. Dual path flow cell technology offers a dynamic absorbance range 40 times greater than conventional detectors, especially for preparative LC applications.
-
Secure compound identification - simultaneously handling two user-defined UV or visible wavelengths increases analyte information and enhances productivity. Peak purity is easily assessed with dual wavelength and peak ratio capabilities.
-
Superb dynamic range – Dual Pathlength flow cell delivers a dynamic absorbance range 40 times greater than conventional detectors.
-
Scale-up easily from analytical to preparative flow rates. Interchangeable, cartridge-style flow cells quickly switch flow rates from a microliter per minute to hundreds of milliliters per minute.
-
OpenLAB CDS ChemStation Edition Software or LC Responder Software provide full system control, data analysis and multiple reporting options plus tools for 21 CFR Part 11 compliance.
-
Stable, precise wavelength settings – Deuterium source is used for the ultraviolet range and a quartz halogen lamp for the visible range.
-
Tool-free maintenance – Routine maintenance is easy with front panel access. Simple thumbscrews secure the lamps and flow cells, no tools are needed.
-
Regulatory compliance – Agilent offers complete IQ, OQ, and PQ documentation packages and contract services to help meet regulatory requirements.
Especificaciones Agilent 325 con doble longitud de onda UV-Vis
Wavelength range |
190 to 900 nm |
Light sources: |
Deuterium for UV and quartz halogen for Vis |
Spectral bandwidth: |
6 nm |
Wavelength precision: |
±0.1 nm |
Wavelength accuracy: |
±1 nm at 260 nm |
Noise, single wavelength: |
±5 x 10-6 AU peak-to-peak at 254 nm, 2 sec response time, 1 mL/min MeOH, 9-mm flow cell |
Noise, dual wavelength: |
±10 x 10-6 AU peak-to-peak at 254/280 nm, |
Noise, dry cell: |
Typically, ±3 x 10-6 AU peak-to-peak at 254 nm, |
Drift: |
Less than 1 mAU/hr |
Linearity, with acetone in water: |
Analytical 1% to 2 AU; preparative1% to 20 AU at 265 nm |
Absorbance range: |
Up to 70 AU with preparative flow cell |
Response time, selectable: |
0.05, 0.5, 1, and 2 sec |
Data rate: |
20 Hz single channel |
Analog outputs: |
Two with selectable absorbance range of 0.001 to 100 AU full scale at 1V |
External: |
Ethernet |
Contact closure outputs: |
Four time-programmable external event relays, one peak sense relay, three synchronization signals (ready out,start out, and fault out) |
Contact closure inputs: |
Lamp off, autozero, three synchronization signals (ready in, start in, and fault in) |
Software: |
Agilent OpenLab ChemStation |
Dual channel ratio calculation: |
Peak detection through OpenLab ChemStation |
Dual pathlength flow cells: |
Preparative flow cell: 9 mm x 1 mm; 4 x 0.25 |
Maximum flow rate preparative flow cell: |
Preparative flow cell: 500 mL/min |
Pressure: |
1,000 psi maximum |