1260 Infinity Multiple Wavelength Detector

Real multiple wavelength detection for ultrafast LC

The 1260 Infinity Multiple Wavelength detector (MWD) will simultaneously detect up to eight compound-specific wavelengths which gives you optimum selectivity over a wavelength range from 190 nm to 950 nm. The improved diode array design greatly reduces detector noise (<± 7 µAU) for precise quantification of trace compounds, regardless of how many signals are recorded. High-speed UV detection with up to 80 Hz data rates keeps pace with the analysis speed of fast LC

  • Higher sensitivity and selectivity - simultaneous detection of up to eight compound-specific wavelengths

  • Low detection limits - low noise front-end electronics and the patented flow cell design delivers very low detection limits thanks to the minimization of short-term noise (< ± 7 μAU).

  • Up to 100% resolution gain in fast LC - using an 80 Hz data acquisition rate.

  • Maximum baseline stability - electronic temperature control (ETC) reduces baseline drift under fluctuating ambient temperature and humidity conditions.

  • Wide linear range - for reliable, simultaneous quantification of primary compounds, by-products and impurities.

  • Programmable slit (1 to 16 nm) for rapid optimization of sensitivity and linearity.

  • Excellent data security and traceability - data is saved to the new data recovery card (DRC) and radio frequency identification (RFID) tags on cells and source lamps improve traceability of data.

  • Automatic wavelength verification by built-in holmium oxide filter.

  • Nine analytical and preparative flow cells provide you with maximum application flexibility and choice.

  • Extensive diagnostics, error detection and display with 1200 Infinity Series Instant Pilot controller and Lab Advisor software.

 

Specifications: 1260 Infinity Multiple Wavelength Detector (G1365C

Detector type

1024-element diode array

Light source

Deuterium and tungsten

Number of signals

8

Maximum sampling rate

80 Hz

Short-term noise

± 0.7 x 10 -5 AU at 254/4 nm and at 750 nm, TC 2 sec.

Drift

< 0.9 x 10-3 AU/hr at 254 nm

Linearity

> 2.0 AU (5 %) at 265 nm

Wavelength range

190-950 nm

Wavelength accuracy

± 1 nm, self-calibration with deuterium lines
verification with holmium oxide filter

Slit width

Programmable: 1, 2, 4, 8, 16 nm

Diode width

< 1 nm

Wavelength bunching

Programmable, 2 - 400 nm, in steps of 1 nm

Flow cells

Standard
13-µL volume, 10 mm-cell path length, 120 bar (1760 psi) pressure maximum
Semi-micro
5-µL volume, 6-mm cell path length, 120 bar (1760 psi) pressure maximum

Micro
2-µL, volume, 3-mm cell path length, 400 bar (5880 psi) pressure maximum
Semi-nano
500-nanoliter volume., 10-mm cell path length, 50 bar (735 psi) pressure maximum
Nano
80-nanoliter volume, 6-mm cell path length, 50 bar (735 psi) pressure maximum
High Pressure
1.7-µL volume, 6-mm cell path length, 400 bar (5880 psi) pressure maximum
Preparative 3 mm
3-mm cell path length, 120 bar (1760 psi) pressure maximum

Preparative 3 mm 
0.3-mm cell path lenght, 20 bar (294 psi) pressure maximum, quartz
Preparative 0.06 mm
0.06-mm cell path length, 20 bar (294 psi) pressure maximum, quartz

Time programmable

Wavelength, polarity, peak width, lamp bandwidth, autobalance, wavelength range, threshold, spectra storage mode

Analog output

Recorder/integrator: 100 mV or 1 V, 2 outputs

Communications

LAN, Controller-area network (CAN),
RS-232C, APG Remote: ready
start, stop and shut-down, signals,

GLP features

Data recovery card to prevent data losses. RFID for electronics records of flow cell and UV lamp conditions (path length, volume, product number, serial number, test passed, usage)

Early maintenance feedback (EMF) for continuous tracking of instrument usage in terms of lamp burn time with user settable limits and feedback messages. Electronic records of maintenance and errors. Verification of wavelength accuracy with built-in holmium oxide filter.

Safety and maintenance

Extensive diagnostics, error detection and display through Agilent Instant Pilot and Agilent Lab Advisor software. Leak detection, safe leak handling, and leak output signal for shutdown of pumping system. Low voltages in major maintenance areas.

Others

Electronic temperature control (ETC) for the complete optical unit

 

Specifications:1260 Infinity Multiple Wavelength Detector VL (G1365D)

Detector type

1024-element diode array

Light source

Deuterium and tungsten

Number of signals

8

Maximum sampling rate

20 Hz

Short-term noise

± 0.7 x 10 -5 AU at 254/4 nm and at 750 nm, TC 2 sec.

Drift

< 0.9 x 10-3 AU/hr at 254 nm

Linearity

> 2.0 AU (5 %) at 265 nm

Wavelength range

190-950 nm

Wavelength accuracy

± 1 nm, self-calibration with deuterium lines
verification with holmium oxide filter

Slit width

Programmable: 1, 2, 4, 8, 16 nm

Diode width

< 1 nm

Wavelength bunching

Programmable, 2 - 400 nm, in steps of 1 nm

Flow cells

Standard
13-µL volume, 10 mm-cell path length, 120 bar (1760 psi) pressure maximum
Semi-micro
5-µL volume, 6-mm cell path length, 120 bar (1760 psi) pressure maximum

Micro
2-µL, volume, 3-mm cell path length, 400 bar (5880 psi) pressure maximum
Semi-nano
500-nanoliter volume., 10-mm cell path length, 50 bar (735 psi) pressure maximum
Nano
80-nanoliter volume, 6-mm cell path length, 50 bar (735 psi) pressure maximum
High Pressure
1.7-µL volume, 6-mm cell path length, 400 bar (5880 psi) pressure maximum
Preparative 3 mm
3-mm cell path length, 120 bar (1760 psi) pressure maximum

Preparative 3 mm 
0.3-mm cell path lenght, 20 bar (294 psi) pressure maximum, quartz
Preparative 0.06 mm
0.06-mm cell path length, 20 bar (294 psi) pressure maximum, quartz

Time programmable

Wavelength, polarity, peak width, lamp bandwidth, autobalance, wavelength range, threshold, spectra storage mode

Analog output

Recorder/integrator: 100 mV or 1 V, 2 outputs

Communications

LAN, Controller-area network (CAN),
RS-232C, APG Remote: ready
start, stop and shut-down
signals,

GLP features

RFID for electronics records of flow cell and UV lamp conditions (path length, volume, product number, serial number, test passed, usage)

Early maintenance feedback (EMF) for continuous tracking of instrument usage in terms of lamp burn time with user settable limits and feedback messages. Electronic records of maintenance and errors. Verification of wavelength accuracy with built-in holmium oxide filter.

Safety and maintenance

Extensive diagnostics, error detection and display through Agilent Instant Pilot and Agilent Lab Advisor software. Leak detection, safe leak handling, leak output signal for shutdown of pumping system. Low voltages in major maintenance areas.

Others

Electronic temperature control (ETC) for the complete optical unit