1260 Infinity Autosampler

The Agilent 1260 Infinity Autosampler is designed for the reliability, safety, and ease-of-use needed for routine pharmaceutical tasks and quality control, as well as for environmental and food analyses. It can house optionally the integrated column compartment for two LC columns with temperature control up to 80 °C as well as a sample cooler for stable temperatures down to 4 °C, all within one module.

  • Reliable injections from 0.1 to 100 µL

  • Capacity for up to 132 vials (2 mL) or up to 36 vials (6 mL)

  • Easy adaption for injection volumes up to 1500 µL for applications ranging from microbore to semipreparative chromatography

  • Includes a needle flush port for rinsing of outside needle to maintain lowest carryover during routine work

  • Integrated column compartment as option or upgrade available, holding two columns up to 30-cm length, and providing heating capacity up to 80 °C for reproducible chromatography data at optimized resolution

  • Integrated sample cooler as option or upgrade available, providing high cooling capacity for all vials loaded down to 4 °C

  • Low internal volume for minimum contribution to a system's total internal volume, which can be even further reduced using "bypass" mode

  • Overlapped injections for increased productivity

  • Injection program available for customizing advanced injections as well as for sample preparation steps upfront injection

 

 

Specifications for 1260 Infinity Standard Autosampler
Valid with  standard 100 µL metering head installed

Injection range

0.1- 100 µL in 0.1 µL increments
Up to 1500 µL with multiple draw (hardware modification required)

Precision

<0.25% RSD from 5-100 µL, <1% from 1-5 µL, variable volume

Sample viscosity range

0.2 - 5 cp

Sample capacity

100 × 2-ml vials in 1 tray
40 × 2-ml vials in ½ tray
15 × 6-ml vials in ½ tray (Agilent vials only)

Injection cycle time

Typically 50 s depending on draw speed and injection volume

Carryover

Typically < 0.1 %, < 0.05 % with external needle cleaning

Pressure operating range

Up to 600 bar (60 MPa or 8702 psi)