1290 Infinity LC Injector HTC/HTS
High throughput with lowest carryover
The Agilent 1290 Infinity LC Injector HTC/HTS is designed specifically to meet the throughput and reliability demands of critical LC/MS applications. The device offers the lowest carryover available with the dynamic load and wash (DLW) principle and enables improved cycle times through the use of faster mechanisms. This highly versatile system is optimized for high throughput injections and perfectly complements the market-leading high speed, resolution and sensitivity of the Agilent 1290 Infinity LC system. Temperature controlled storage conserves sample integrity and a variety of software options are available for different applications.
-
Near zero carryover provided by the innovative dynamic load and wash principle (DLW).
-
Improved cycle time created by faster robotics and washing mechanisms.
-
Proprietary 1200 bar valve for more robust UHPLC analysis — available exclusively from Agilent.
-
High sample capacity for up to 24 cooled well plates.
-
High level of compatibility with easy control and data handling through Mass Hunter, OpenLab Chemstation and OpenLab EZChrom software.
-
Temperature-controlled sample storage that keeps a variety of samples cool and stable preventing degradation.
System Type |
|
Syringe size |
100 µL |
Loop sizes for DLW and maximum pressure rating |
2 µL (standard) 1200 bar max., |
LC injector |
1 electrically actuated fast switching 6-port valve from Rheodyne (1200 bar) |
Injection volumes |
Up to 100 µL with standard setup, up to 5 ml with additional hardware |
HTC sample capacity |
Up to 24 shallow well-plates (96 or 384) |
Injection precision |
<0. 5% with 100 ul syringe and 2 ul loop, overfill 7ul |
HTS sample capacity |
Up to 24 shallow well-plates (96 or 384) |
Needle, syringe valve cleaning |
Dynamic load and wash (DLW) principle: two different solvents |
Carryover |
Typically < 0.004% with DLW Method: Chlorohexidine, 0.6 mg/L dissolved in H2O : ACN = 90 : 10, 0.1% TFA |
Local control |
Handheld control panel with 4 function keys, graphical LCD display |
Electronic control |
2 RS232C ports |
Thermal control |
4-40C (settable) |
System Type |
XYZ Robot With Syringe-Only Concept, No Tubing In Sample Path |
Syringe size |
100 µL |
Loop sizes for DLW and maximum pressure rating |
2 µL (standard) 1200 bar max., |
LC injector |
1 electrically actuated fast switching 6-port valve from Rheodyne (1200 bar) |
Injection volumes |
Up to 100 µL with standard setup, up to 5 ml with additional hardware |
HTC sample capacity |
Up to 24 shallow well-plates (96 or 384) |
Injection precision |
<0. 5% with 100 ul syringe and 2 ul loop, overfill 7ul |
HTS sample capacity |
Up to 24 shallow well-plates (96 or 384) |