HG ST BASIC

The HydroGen series HG ST BASICis able to generate up to 350 cc/min of Hydrogen with purity better than 99.9999% and pressure up to 10 bar (145 psi). The HG ST BASIC includes a new long life multilayer electrolytic cell with polymeric membrane (PEM) using distilled water. No acid or alkaline solutions are used. 

The innovative drying system ensures the maximum grade of hydrogen purity, protects the GC thanks to its innovative humidity safe guard, HG ST BASIC is completely maintenance-free and allows continuous operation, 24 hours a day. The on-board CPU automatically check for internal leaks and constantly control the operating parameters to guarantee full safety.

Up to 20 units can connected in parallel mode using an external control box. The touch-screen LCD interface provides simple and user-friendly management of all functions on the unit.

Main Applications

  •  GC-FID/ GC-NPD/ GC-FPD
  • THA
  • Gas sensing
Main Advantages
  • Quick installation
  • Easy use
  • Hydrogen flow rate up to 350 cc/min
  • ZeroAir flow rate up to 2000 cc/min
  • Pressure up to 10 bar (145 psi)
  • Hydrogen purity: 99.9999%
  • THC: < 0.1 ppm
  • RS 485 and USB as standard
  • LAN optional
  • Dryer with no maintenance
  • Reduced foot print
  • High pressure GLS electronically controlled (patented)

Models: HG ST BASIC

100

160

250

350

General data

Electrolytic cell

PEM technology

Drying system

Static dryer

H2 purity

>99.99999%*

Outlet pressure

7 bars (100psi)/ 10 bars (145 psi) (optional)

H2 flow-rate cc/min (Max)

100

160

250

350

Dimensions (WxHxD)

9.5 x 44 x 48 cm

Net weight (no water in tank)

18 kg

Communication

USB/ RS-485

Standard

LAN

Optional

Software functions

Parallel mode

Ready on board

Automatic tank refill

Standard

Water

Quality Required

Deionized, ASTM II, <0.1uS

Supply pressure (min, Max)

-0.2 bars (1.4 psi), 0.5 bar (7.25 psi)

Internal tank capacity

0.3 l

External tank capacity

5 or 10 L (optional)

Electrical data

Type of connection

IEC 320-C13

Power supply voltage

100-240 Vac 50/60 Hz

Installed power (Max)

350W

Fuse rating (5×20mm)

4A 250VAC

Connections

Hydrogen outlet

1/8” compression fitting

Water (inlet, drain)

Quick release push-in fitting

ZeroAir section (optional)

Inlet pressure (Max)

8 bars (116 psi)

THC (Inlet Air)

100 ppm

Outlet flow rate (Max)

2000 cc/min

Outlet pressure (max)

Inlet pressure – 0.5 bars (8 psi) at maximum flow

THC (ZeroAir Outlet)

<0.1 ppm

Inlet, Outlet ports

1/8” BSPP female

 * Referred to O2, Dew point <-25°C

Ordering codes

Gas Generators
6920.05.010 HG ST BASIC 100
6920.05.016 HG ST BASIC 160
6920.05.025 HG ST BASIC 250
6920.05.035 HG ST BASIC 350
Accessories (optionals)
6930.00.120 ZA 2U (catalyst)
Consumable
6930.00.050 Deionisation Cartridge