ELSD 7 Angles SLD7000

The SECcurity SLD7000 MALLS (MALS) is a 7-angle static light scattering detector for the measurement of absolute molecular weight, size, and conformation of macromolecules in solution. The revolutionary design with a cylindrical, self-cleaning, and index-matching cell avoids common pitfalls in GPC/SEC-light scattering detection. Modern ultra-sensitive CCD detection results in high sensitivity, even at low concentration and molar masses.

The small footprint allows integration into any GPC/SEC or FFF system. The SLD7000 can be used in batch mode (off-line) without modifications. In combination with an RI and a viscosity detector the light scattering detector can be used for triple detection and triple plus detection.

 

 Specifications

 Number of angles

 7 plus additional reference, simultaneous detection 

 Detection angles/  scattering angles

 35, 50, 75, 90 105, 130, 145

 No difference detection angle/scattering angle due to index-matching cell

 Cell

 PEEK

 Cell volumne

 100 µL nominal

 Scattering volume

 20 nL nominal

 Cell cleaning option  

 Self cleaning cell with vertical flow path

 Light source

 35 mW, 635 nm, vertically polarized temperature stabilized with precision power control

 Detector

 Fibre optics with low numerical aperture, integral to cell CCD with ultra-high sensitivity and spatial uniformity

 Control

 Integrated, dedicated, powerful microcontroller

 Input

 4 analog ins with 16 and 24 bit resolution for any concentration detector or viscometer autosampler/manual injector  digital inject in

 Options

  • temperature control: P/N 403-0005
  • metal-free detector SLD7000B: P/N 403-0010
  • other cell materials on request
  • fluorescence filter: P/N 403-001
  • Software for static light scattering measurements in batch mode P/N 403-0004

 GLP/Compliance

 Fan on/off with built-in temperature control