Agilent 7697A
The Agilent 7697A Headspace Sampler, in both 111 and 12 vial capacity configurations, represents the state-of-the-art in headspace gas chromatography. It is based on the architecture of the market leading 7890 Series GC and 7693A ALS line of products.
Independent control of vial pressurization and column head pressure provides increased precision. Barometric pressure compensation ensures the increased precision is consistent from day to day and across multiple labs. Perform method conversion from any headspace method to 7697A without any rework.
Learn more by watching the 7697A Headspace Sampler Video.
- The 7697A Headspace Sampler 12 Vial Capacity configuration is meant for academic labs and those with small sample loads
- Standard 111 sample configuration with 3 priority samples
- Exchangeable trays during a sequence for increased sample capacity
- Hydrogen carrier and auto Sleep/Wake modes conserve valuable laboratory resources.
- Independent control of vial pressurization and column head pressure provides increased precision by ensuring no back pressure during sample transfer.
- Automatic vial leak test checks each vial during pressurization – without time consuming calibration - to ensure that every sample is properly capped.
- Sample Bar Code Reader with data transfer capabilities allows you to track samples before they enter the oven.
- Multiple headspace concentrations (MHC) with up to 100 extractions per GC start allows for a more complete sampling and increased sensitivity
- Multiple headspace extractions (MHE) with up to 100 extractions per vial
- Supports 10 mL, 20 mL, or 22 mL vials without an adapter
- Method development software transfers methods from any headspace system or quickly designs a method from scratch using just a few known parameters.
- Samples depressurized before returning to tray
Cromatographic Performance | |
Reproducibility | ≤ 1,5% RSD |
Reproducibility with tray | ≤ 1% RSD |
Basic System Capacity | |
Positions | 12 vial in a single position |
Complet System Capacity | |
Positions | 111 vial |
Trays | 3 trays of 36 vial |
Priority positions | Up to 3 trays |
Cooling | Available for 108 vial |
Bar code reader | Option integrated |
Vial | 10, 20 or 22 ml |
Control System | |
Purge | Automatized after every sample |
LED Indicators | Not ready / Run / Sleep / Service Due / Tray Park |
Methods | Store up to 32 user-defined methods |
Sequences | Store up to 9 |
Modes of Operation | |
HE | Headspace Extraction |
MHE | Multiple Headspace Extraction |
MHC | Multiple Headspace Concentration |
Thermal Control | |
Temperature zones | All temperatures zones (oven, valve and loop, transfer line) have set point increments |
Set point increments | 1ºC with 0.1 ºC resolution |
Pneumatic Control(EPC) | |
External pressure | Compensation for barometric pressure |
Increments pressure | Adjusted by 0,001 psi |
Pressure range | 0,000 to 75,000 psi |
Pressure sensors | Repeatability < +/- 0,05 psi, temperature coefficient < +/- 0,01 psi/ºC |
Timing Control | |
Vial equilibration time | 0 to 999,990 min in 0,001 min increments |
Injection duration | 0 to 999,990 min in 0,001 min increments |
GC cycle time | 0 to 999,99 min in 0,001 min increments |
Purge time | 0 to 999,99 min in 0,001 min increments |
Sample Pathway | |
Inert | Yes |
Loop options | 0,025/0,050/0,100/0,500 / 1 / 3ml in inert material |
Columns compatible with the system | 50 to 530 µm |
Transfer line | 1 m with tubing type of 0,25, 0,2, 0,53 or 0,67 mm ID |
Environmental Conditions | |
Temperature in operation | 10 to 40 ºC |
Temperature in storage | (-40 to 70 ºC) |
Other Specifications | |
Weight | 38 Kg and 46 Kg with tray |
Dimensions | 606 mm (H) x 509 mm (W) x 636 mm (D) and 800 mm (H) x 509 mm (W) x 636 mm (D) with tray |