1260 Infinity Standard Autosampler
The Agilent 1260 Infinity Standard Autosampler is designed for the reliability, safety, and ease-of-use needed in pharmaceutical routine tasks and quality control, as well as for environmental and food analyses. It can be upgraded from non-thermostatted version to thermostatted version by adding the Agilent 1290 Infinity Thermostat.
-
Reliable injections from 0.1 µL to 100 µL.
-
Easy adaption for injection volumes up to 1500 µL for applications ranging from microbore to semipreparative chromatography.
-
Low internal volume of 300 µL for the minimum contribution to a system's total internal volume can be even further reduced using "by-pass" mode.
-
Overlapped injections for increased productivity.
-
Peltier temperature control from 4 to 40 °C for up to 100 2-mL vials containing thermally labile samples (upgrade from non-thermostatted to thermostatted version possible).
Specifications for 1260 Infinity Standard Autosampler |
|
Injection range |
0.1- 100 µL in 0.1 µL increments |
Precision |
<0.25% RSD from 5-100 µL, <1% from 1-5 µL, variable volume |
Sample viscosity range |
0.2 - 5 cp |
Sample capacity |
100 × 2-ml vials in 1 tray |
Injection cycle time |
Typically 50 s depending on draw speed and injection volume |
Carryover |
Typically < 0.1 %, < 0.05 % with external needle cleaning |
Pressure operating range |
Up to 600 bar (60 MPa or 8702 psi) |
Specifications for 1260 Infinity Standard Autosampler |
|
Injection range |
0.1 - 900 µL in 0.1 µL increments (recommended 1 µL increments) |
Precision |
Typically < 0.5 % RSD of peak areas from 5 - 2000 µL |
Sample viscosity range |
0.2 - 5 cp |
Sample capacity |
100 × 2-ml vials in 1 tray |
Injection cycle time |
Typically 50 s depending on draw speed and injection volume |
Carryover |
Typically < 0.1 %, < 0.05 % with external needle cleaning |
Pressure operating range |
Up to 200 bar (20 MPa or 2901 psi) |