1260 Infinity Standard Autosampler

The Agilent 1260 Infinity Standard Autosampler is designed for the reliability, safety, and ease-of-use needed in pharmaceutical routine tasks and quality control, as well as for environmental and food analyses. It can be upgraded from non-thermostatted version to thermostatted version by adding the Agilent 1290 Infinity Thermostat.

  • Reliable injections from 0.1 µL to 100 µL.

  • Easy adaption for injection volumes up to 1500 µL for applications ranging from microbore to semipreparative chromatography.

  • Low internal volume of 300 µL for the minimum contribution to a system's total internal volume can be even further reduced using "by-pass" mode.

  • Overlapped injections for increased productivity.

  • Peltier temperature control from 4 to 40 °C for up to 100 2-mL vials containing thermally labile samples (upgrade from non-thermostatted to thermostatted version possible).

 

Specifications for 1260 Infinity Standard Autosampler
Valid with  standard 100 µL metering head installed

Injection range

0.1- 100 µL in 0.1 µL increments
Up to 1500 µL with multiple draw (hardware modification required)

Precision

<0.25% RSD from 5-100 µL, <1% from 1-5 µL, variable volume

Sample viscosity range

0.2 - 5 cp

Sample capacity

100 × 2-ml vials in 1 tray
40 × 2-ml vials in ½ tray
15 × 6-ml vials in ½ tray (Agilent vials only)

Injection cycle time

Typically 50 s depending on draw speed and injection volume

Carryover

Typically < 0.1 %, < 0.05 % with external needle cleaning

Pressure operating range

Up to 600 bar (60 MPa or 8702 psi)

 

Specifications for 1260 Infinity Standard Autosampler
Valid with standard 900 µL metering head installed

Injection range

0.1 - 900 µL in 0.1 µL increments (recommended 1 µL increments)
Up to 1800 µL with multiple draw (hardware modification required)

Precision

Typically < 0.5 % RSD of peak areas from 5 - 2000 µL
Typically < 1 % RSD of peak areas from 2000 - 5000 µL
Typically < 3 % RSD of peak areas from 1 - 5 µL
Variable volume

Sample viscosity range

0.2 - 5 cp

Sample capacity

100 × 2-ml vials in 1 tray
40 × 2-ml vials in ½ tray
15 × 6-ml vials in ½ tray (Agilent vials only)

Injection cycle time

Typically 50 s depending on draw speed and injection volume

Carryover

Typically < 0.1 %, < 0.05 % with external needle cleaning

Pressure operating range

Up to 200 bar (20 MPa or 2901 psi)