1260 Infinity Quaternary Pump

This versatile pump now has an extended power range, delivering pressures up to 600 bars. At these high pressures you can use smaller particle size columns and get higher resolution and faster separations. It maintains virtually pulse-free, well mixed and stable solvent flows. Its dual floating, precise servo-controlled pistons adjust the stroke volume according to your chosen flow rate. Offering access to up to four solvents, the 1260 Infinity Quaternary Pump provides the greatest flexibility in automated solvent blending and is recommended for a wide range of research and routine applications (especially in the food, environmental and pharmaceutical sectors) as well as for method development.

  • Higher resolution, faster separations - the maximum pressure of 600 bars allows you to use smaller particle size columns, longer columns and/or higher viscous solvents as an alternative to acetonitrile.

  • Analytical and semi-preparative purifications - the wide flow range (up to 10 ml/min) perfectly supports standard and semi-preparative applications. The perfect choice for 4.6 and 3.0 mm id columns.

  • Quicker experiments - the pump offers convenient access to up to four solvents for isocratic or gradient analysis which can really speed up method development, preparation of mobile phases and system flushing.

  • Easy, fast and secure transfer of methods from HPLC to sub-2-micron columns.

  • Integrated degasser

  • Longevity and low operating costs - robust materials (stainless steel, titanium, gold, ruby, sapphire, ceramics, PEEK and PTFE) withstand the most demanding applications and keep your pump functioning for years. Our optional seal wash for high-salt mobile phases keeps maintenance to a minimum and reduces operating costs

 

 

 

Specifications Agilent 1260 Infinity Quaternary Pump (G1311B) and VL version (G1311C)

Hydraulic system

Dual piston in series pump with servo-controlled variable stroke drive, floating pistons

Settable flow range

Set points 0.001 - 10 mL/min, in 0.001 mL/min increments

Flow range

0.2 - 10.0 mL/min

Flow precision

<= 0.07 % RSD or <= 0.02 min SD, whatever is greater; based on retention time at constant room temperature

Flow accuracy

± 1 % or 10 µL/min whatever is greater; pumping degassed H2O at 10 MPa
(100 bar)

Pressure operating range

1260 Infinity Quaternary pump:
Operating range up to 60 MPa (600 bar, 8700 psi) up to 5 mL/min
Operating range up to 20 MPa (200 bar, 2950 psi) up to 10 mL/min

1260 Infinity Quaternary pump VL:
Operating range up to 40 MPa (400 bar, 5880 psi) up to 5 mL/min
Operating range up to 20 MPa (200 bar, 2950 psi) up to 10 mL/min

Pressure pulsation

< 2 % amplitude (typically < 1.3 %) or < 0.3 MPa (3 bar), whatever is greater, at
1 mL/min isopropanol, at all pressures > 1 MPa (10 bar, 147 psi)

Compressibility compensation

User selectable, based on mobile phase compressibility

Recommended pH range

1.0 - 12.5, solvents with pH < 2.3 should not contain acids which attack stainless steel

Gradient formation

Low pressure quaternary mixing/gradient capability using high-speed proportioning valve

Delay volume

600 – 900 μL, dependent on back pressure;  measured with water at 1 mL/min (water/caffeine tracer)

Composition range

0 - 95 % or 5 - 100 %, user selectable

Composition precision

< 0.2 % RSD or < 0.04 min SD, whatever is greater, at 1 mL/min; based on retention time at constant room temperature

Integrated degassing unit

Number of channels: 4
Internal volume per channel: 1.5 mL
Materials in contact with solvent: TFE/PDD Copolymer, FEP, PEEK, PPS

Control

Agilent control software (e.g. ChemStation, EZChrom, OL, MassHunter)

Local control

Agilent Instant Pilot

Communications

Controller-area network (CAN), RS-232C, APG Remote: ready, start, stop and shut-down signals, LAN optional

Safety and maintenance

Extensive diagnostics, error detection and display through Agilent LabAdvisor, leak detection, safe leak handling, leak output signal for shutdown of the pumping system. Low voltage in major maintenance areas.

GLP features

Early maintenance feedback (EMF) for continuous tracking of instrument usage in terms of seal wear and volume of pumped mobile phase with pre-defined and user settable limits and feedback messages. Electronic records of maintenance and errors

Housing