1260 Infinity NanoFlow LC System
Nanoflow LC is widely used for high resolution separations and high sensitivity MS detection. The 1260 Infinity Nanoflow LC system incorporates Agilent's unique Electronic Flow Control (EFC) with active feedback and real-time flow adjustment. EFC delivers a constant flow to the column that is independent of back pressure in the system. Combined with Agilent's capillary and nanoflow pumps, EFC is the gold standard for reliable and robust nanoflow performance. The 1260 Infinity Nanoflow LC is extremely easy to use; you simply select your nanoflow rate and the system then works to deliver this precise flow alongside the gradient performance required for reproducible retention times. Precise flow rates support stable ion generation to ensure you obtain high sensitivity nanoflow LC/MS performance.
-
Unsurpassed low-flow performance and stability - the nanoflow pump features proven Electronic Flow Control (EFC) with active feedback for unprecedented stability at flow rates from 100 nl/min to 1 µl/min.
-
Reproducible retention times - EFC with real-time flow control delivers a constant nanoflow that is independent of system back pressure, so retention times are highly reproducible and nanospray operation is extremely stable.
-
Flexibility for non-nano applications - in addition to nanoflow operations, the system offers full gradient capability for up to 2.5 ml/min in 'normal mode' adding to the system's versatility.
-
Maximum separation efficiency - the switching valves, fittings, connectors and PEEK-coated fused silica tubing have all been carefully selected to minimize dead volumes.
-
Automated sample cleanup, enrichment and 2D HPLC with conventional nanocolumns.
-
Conversion to 2D HPLC by adding another column (e.g. ion exchange column).
-
Easy-to-operate nanospray LC/MS with the innovative and fully integrated HPLC-Chip/MS system
-
Key component of the Agilent Nanospray Protein ID solution and the HPLC-Chip/MS Protein ID solution.
-
Fully compatible with all Agilent mass spectrometers and with most third party ESI, nanospray or MALDI MS systems. Versatile software to control several platforms
Agilent 1260 Infinity Nanoflow Pump |
|
Hydraulic system |
Two dual piston in series, with proprietary servo-controlled variable stroke drive, floating piston, active inlet valve, solvent selection valve and electronic flow control for flow rates from 0.1 to 1 µL/min |
Settable column flow range |
0.01 – 4 µL/min, |
Recommended column flow range |
0.1 – 1 µL/min |
Optimum composition range |
1 to 99% or 5 µL/min per channel (primary flow), whatever is greater |
Composition precision |
< 0.2 % SD, at 500 nl/min (default settings) |
Delay volume |
Typically 300 nL from the electronic flow control to the pump outlet for flow rates up to 4 µL/min. |
Pressure range |
20 to 400 bar (5880 psi) system pressure |
Compressibility compensation |
User-selectable, based on mobile phase compressibility |
Recommended pH range |
1.0 – 8.5, solvents with pH < 2.3 should not contain acids which attack stainless steel. Upper pH range is limited by fused silica capillaries. |
Control and data evaluation |
Agilent ChemStation for LC and Agilent MassHunter for LC/MS |
Analog output |
For pressure monitoring, 2 mV/bar, one output |
Communications |
Controller-area network (CAN), GPIB, RS-232C, APG Remote: ready, start, stop and shut-down signals, LAN optional |